Female breast radiation exposure during thorax multidetector computed tomography and the effectiveness of bismuth breast shield to reduce breast radiation dose
Related Experiment Video
Updated: Jul 17, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Mehmet Halit Yilmaz1, Sait Albayram, Doğan Yaşar
1Department of Radiology, Cerrahpasa Medical Faculty, Istanbul University, Turkey. drmhyilmaz@yahoo.com
Objective:
The purpose of our study was to determine the breast radiation dose when performing routine thoracic multidetector computed tomography (MDCT). We also evaluated dose reduction and the effect on image quality of using a bismuth breast shield when performing thoracic MDCT.
Material And Methods:
The dose reduction achievable by shielding the adult (18 years or older) female breasts was studied in 50 women who underwent routine thoracic MDCT. All examinations were performed with a 16-MDCT scanner (Sensation Cardiac 16; Siemens Medical Solutions). To compare the shielded/unshielded breast dose, the examination was performed with (right breast) and without (left breast) breast shielding in all patients. With this technique, the superficial breast doses were calculated. To determine the average glandular breast radiation dose, we imaged an anthropomorphic dosimetric phantom into which calibrated dosimeters were placed to measure the dose to breast. The phantom was imaged using the same protocol. Radiation doses to the breasts with and without the breast shielding were measured and compared using the Student t test.
Results:
In the qualitative evaluation of the MDCT scans, all were considered to be of diagnostic quality. We did not see any differences in quality between the shielded and unshielded lung. The mean radiation doses to the breasts with the shield and to those without the shield were 8.6 +/- 2.33 versus 14.46 +/- 3.94 mGy, respectively. The breast shield enabled a 40.53% decrease in radiation dose to the breast. The difference between the dose received by the breasts with and that received by the breasts without bismuth shielding was significant, with a P value of less than 0.001.
Conclusions:
Bismuth in-plane shielding for routine thoracic MDCT decreased radiation dose to the breast without qualitative changes in image quality. The other radiosensitive superficial organs (eg, testes and thyroid gland) specifically must be protected with shielding.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...