Related Experiment Video
Updated: Jul 17, 2026

Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Personal dosimetry in terms of HP(3): Monte Carlo and experimental studies
P Ferrari1, G Gualdrini, R Bedogni
1ENEA-ION-IRP Radiation Protection Institute ENEA-ION-IRP Radiation Protection Institute, Via dei Colli 16, 40136, Bologna (BO), Italy. paolo.ferrari@bologna.enea.it
Abstract:
Hp(3) has been defined as the operational quantity for eye lens dosimetry. Hp(3)/ka conversion coefficients were evaluated at the GSF (Germany) in a 30x30x15 cm3 4-elements ICRU slab phantom for various energies and incident angles through Monte Carlo. The ISO report 12,794 suggests to employ a PMMA water filled phantom, of the same dimensions, for dosemeter calibration in terms of Hp(3). The present paper briefly summarises the main aspects of a study carried out at ENEA-Radiation Protection Institute (Bologna, Italy) to provide practical procedures for the calibration of dosemeters in terms of Hp(3). Tabulations of a new set conversion coefficients and air kerma backscatter factors are provided as a function of energy and incident angle. The paper demonstrates that a more accurate approach to the dosimetric assessment in terms of Hp(3) could be rather simply introduced employing a reduced phantom.