A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
A new technique for assessing arterial pressure wave forms and central pressure with tissue Doppler
Brian A Haluska1, Leanne Jeffriess, Phillip M Mottram
1University of Queensland Department of Medicine, Princess Alexandra Hospital, Ipswich Rd, Brisbane Q4102, Australia. b.haluska@uq.edu.au
Background:
Non-invasive assessment of arterial pressure wave forms using applanation tonometry of the radial or carotid arteries can be technically challenging and has not found wide clinical application. 2D imaging of the common carotid arteries is routinely used and we sought to determine whether arterial waveform measurements could be derived from tissue Doppler imaging (TDI) of the carotid artery.
Methods:
We studied 91 subjects (52 men, age 52 +/- 14 years) with and without cardiovascular disease. Tonometry was performed on the carotid artery simultaneously with pulsed wave Doppler of the LVOT and acquired digitally. Longitudinal 2D images of the common carotid artery with and without TDI were also acquired digitally and both TDI and tonometry were calibrated using mean and diastolic cuff pressure and analysed off line.
Results:
Correlation between central pressure by TDI and tonometry was excellent for maximum pressure (r = 0.97, p < 0.0001). The mean differences between central pressures derived by TDI and tonometry were minimal (systolic 5.36 +/- 5.5 mmHg; diastolic 1.2 +/- 1.2 mmHg).
Conclusion:
Imaging of the common carotid artery motion with tissue Doppler may permit acquisition of a waveform analogous to that from tonometry. This method may simplify estimation of central arterial pressure and calculation of total arterial compliance.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.