Effect of albumin concentration on serum cortisol measured by the Bayer Advia Centaur assay
1St Thomas' Hospital, Lambeth Palace Road, London SE1 7EH, UK. sophie.barnes@gstt.nhs.uk
Insights
Albumin levels significantly impact cortisol test accuracy. The Bayer Advia Centaur assay shows under-recovery with low albumin, affecting adrenal insufficiency diagnosis in critically ill patients.
Area of Science:
- Clinical Chemistry
- Endocrinology
- Laboratory Medicine
Background:
- Discrepancies noted between Bayer Advia Centaur and Roche Elecsys cortisol assays.
- Assays were performed on samples from endocrine outpatients and intensive therapy unit patients.
Purpose of the Study:
- To investigate the influence of albumin concentration on the performance of two cortisol assays.
- To identify factors contributing to observed differences between the Bayer Advia Centaur and Roche Elecsys cortisol methods.
Main Methods:
- Retrospective analysis of patient samples with recorded albumin concentrations.
- Performed recovery experiments using samples with varying albumin and cortisol levels.
Main Results:
- Albumin concentration explained nearly 50% of the variability in results between the two cortisol assays.
- The Bayer Advia Centaur assay demonstrated significant under-recovery when albumin concentrations fell below 24 g/L.
Conclusions:
- Low albumin concentrations critically affect the accuracy of the Bayer Advia Centaur cortisol assay.
- This under-recovery poses a diagnostic challenge for adrenal insufficiency in critically ill patients with hypoalbuminemia.
Introduction:
A poor comparison was seen between the Bayer Advia Centaur and the Roche Elecsys cortisol methods using samples sent to the laboratory for cortisol analysis. They were predominantly from patients attending the Endocrine outpatients clinic or patients in the intensive therapy unit. We investigated the effect of albumin concentration on the performance of both assays.
Methods:
For the samples used in the patient comparison, the laboratory database was searched for any concomitant albumin concentrations (Beckman LX20 BCP method). A series of recovery experiments were performed combining samples with normal albumin and undetectable cortisol concentrations with samples with detectable cortisol but low albumin concentrations.
Results:
Albumin concentration accounts for almost 50% of the variation in the difference between the two methods. The Bayer Advia Centaur method shows significant under-recovery at albumin concentrations lower than 24 g/L.
Conclusion:
This under-recovery will significantly affect the diagnosis of adrenal insufficiency in critically ill patients with low albumin concentrations.