Related Experiment Video
Updated: Jul 17, 2026

Insertion of Flexible Neural Probes Using Rigid Stiffeners Attached with Biodissolvable Adhesive
Published on: September 27, 2013
Improving mechanical stiffness of coated benzocyclobutene (BCB) based neural implant
Amarjit Singh1, Haixin Zhu, Jiping He
1Center for Neural Interface Design, Arizona State University, Tempe, AZ, USA.
Abstract:
We briefly report recent results of a simple alternate method to improve mechanical stiffness of BCB polymer neural implant for surgical insertion into brain tissue, which uses coatings dissolvable in bio-fluids. We have studied three different coating materials such as thermo-reversible gel Poloxamer 407, glucose (C6H12O6) and regular table sugar that were applied by dip coating onto the implant surface. The preliminary results of this study have shown that coating BCB probes with Poloxamer 407 polymer, a thermo-reversible gel, or table sugar significantly improves the buckling strength. However, the table sugar coating provides the greatest increase in stiffness, which is sufficient to penetrate both the preserved and live brain tissues without buckling.