Brain hemorrhage: evaluation with fast spin-echo and conventional dual spin-echo images
K M Jones1, R V Mulkern, M T Mantello
1Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA 02115.
Radiology
|January 1, 1992
Summary
Spin-echo (SE) and fast spin-echo (FSE) imaging show comparable contrast for central nervous system hemorrhage. Lesion conspicuity and signal intensity differences between SE and FSE sequences were minimal for blood products.
Area of Science:
- Radiology
- Neuroimaging
- Magnetic Resonance Imaging
Background:
- Central nervous system (CNS) hemorrhage evaluation relies on magnetic resonance imaging (MRI).
- Spin-echo (SE) and fast spin-echo (FSE) are common MRI sequences with differing T2-weighting properties.
- Understanding signal intensity differences in hemorrhagic lesions is crucial for accurate diagnosis.
Purpose of the Study:
- To compare signal intensity and contrast of blood products in CNS hemorrhage using SE and FSE sequences.
- To assess the impact of magnetic susceptibility effects on image signal intensity.
- To determine if FSE offers advantages over SE for visualizing hemorrhagic lesions.
Main Methods:
- Evaluated signal intensity in 15 patients with CNS hemorrhage.
- Measured signal intensity in regions of interest including hemorrhage, iron-containing nuclei, white matter, scalp fat, and noise.
- Compared measurements between SE and FSE sequences along the phase-encoding direction.
Main Results:
- Hemosiderin, deoxyhemoglobin, and iron-containing nuclei showed slightly higher signal intensity on FSE, but not significantly.
- Methemoglobin signal intensity was similar between SE and FSE.
- Scalp fat signal intensity was higher on FSE images.
- Signal-to-noise ratios were lower with FSE compared to SE imaging.
- Contrast and lesion conspicuity were comparable between the two sequences.
Conclusions:
- Paramagnetic blood products may exhibit marginally higher signal intensity on FSE images.
- SE and FSE sequences provide comparable contrast and lesion conspicuity for CNS hemorrhage.
- FSE imaging does not significantly alter the appearance of hemorrhagic lesions compared to SE imaging.