Related Experiment Video
Updated: Jul 17, 2026

Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Does repeated disinfection of the e-Flow rapid nebulizer affect in vitro performance?
Lasha Bakuridze1, Véronique Andrieu, Christian Dupont
1Laboratoire de Pharmacie Galénique, Faculté de Pharmacie, Marseille, France.
Abstract:
Repeated disinfection of a nebulizer may modify its performance. During 60 cycles of disinfection with the Nuk steam sterilizer (120 nebulizations), the median volume diameter of tobramycin or colistin, nebulization time and various parameters correlated with the membrane weight of the e-Flow rapid are not modified. Results are similar when tap or purified water is used for washing and disinfection.
More Related Videos
06:27Automated Hospital Room Disinfection Utilizing a Novel Aerosolized Hydrogen Peroxide Microdroplet Disbursing Technology
Published on: February 24, 2026
09:02An Experimental Model to Study Tuberculosis-Malaria Coinfection upon Natural Transmission of Mycobacterium tuberculosis and Plasmodium berghei
Published on: February 17, 2014