Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
An Egsnrc investigation of the P(TP) correction factor for ion chambers in kilovoltage X rays
Daniel J La Russa1, D W O Rogers
1Carleton Laboratory for Radiotherapy Physics, Ottawa-Carleton Institute for Physics, Carleton University, Ottawa, ON, KIS 5B6, Canada. dlarussa@physics.carleton.ca
Abstract:
As part of the standard practice for obtaining consistent ion chamber measurements with cavities open to the surrounding atmosphere, the raw measured response is corrected to the response at a reference temperature and pressure using the standard temperature-pressure correction factor (P(TP)). In this study, the EGSnrc Monte Carlo code was used to investigate the validity of the P(TP) correction factor for kilovoltage x rays incident on various geometrically distinct ion chambers. The calculated P(TP)-corrected chamber response deviated by over 2% relative to expected values for a 40 kV spectrum incident on a graphite thimble chamber at an air density typical of Mexico City. The relative deviation from the expected response was much worse for a large spherical graphite chamber, exceeding 16% at an air density of 0.6 kg/m3 (approximately 0.5 atm at 22 degrees C) for the same beam energy. The breakdown of the P(TP) correction factor was also observed for a 26 kV mammography spectrum incident on two mammography chambers. For 60Co beams, the P(TP) correction factor behaved as expected. For day-to-day variations in pressure, only a negligible of the P(TP) correction factor was observed with low x-ray energies. Factors contributing to the breakdown of the P(TP) correction factor at low x-ray energies and large pressure variations, such as the range of electrons, the material of the wall, the chamber dimensions and air-photon interactions, are discussed in depth.
Related Concept Videos
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Atomic Emission Spectroscopy: Lab
Biological Effects of Radiation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Overview