Common mode noise cancellation for electrically non-contact ECG measurement system on a chair
Ko Keun Kim1, Yong Kyu Lim, Kwang Suk Park
1Interdisciplinary Program in Medical and Biological Engineering, Seoul National University, Republic of Korea.
Abstract:
Electrically non-contact ECG measurement system on a chair can be applied to a number of various fields for continuous health monitoring in daily life. However, the body is floated electrically for this system due to the capacitive electrodes and the floated body is very sensitive to the external noises or motion artifacts which affect the measurement system as the common mode noise. In this paper, the Driven-Seat-Ground circuit similar to the Driven-Right-Leg circuit is proposed to reduce the common mode noise. The analysis of this equivalent circuit is performed and the output signal waveforms are compared between with Driven-Seat-Ground and with capacitive ground. As the results, the Driven-Seat-Ground circuit improves significantly the properties of the fully capacitive ECG measurement system as the negative feedback.