Evaluation of dose exposure in 64-slice CT colonography
O Luz1, M Buchgeister, M Klabunde
1Department of Diagnostic Radiology, University Hospital Tübingen, Hoppe-Seyler-Str 3, 72076, Tübingen, Germany. oliver.luz@med.uni-tuebingen.de
Abstract:
The radiation exposure of four different 64-slice MDCT-colonography (CTC) protocols was evaluated using an Alderson-Rando phantom. Protocols using 30 mAs (collimation 20 x 1.2 mm), 50 mAs (collimation 20 x 1.2 and 64 x 0.6 mm) and 80 mAs (20 x 1.2 mm) representing screening low-dose, routine, narrow collimation and oncologic staging setups were measured with an Alderson-Rando phantom (Alderson Research Laboratories Inc.). Scans were performed on a 64-row MDCT (SOMATOM Sensation 64, Siemens) simulating the prone and supine positions with a constant voltage of 120 kV. Dose values (male/female) were 2.5/2.9, 3.8/4.2, 4.2/4.5 and 5.7/6.4 mSv for 30, 50 (20 x 1.2 and 64 x 0.6 mm) and 80 mAs, respectively. Measurements showed an elevated dose for females (11.5% mean; compared to males). Use of narrow collimation combined with 50 mAs resulted in a small increase of dose exposure of 10.5 (male) and 7.1% (female). Gonad doses ranged from 0.9 to 2.6 mSv (male) and from 1.5 to 3.5 mSv (female). In all protocols, the stomach wall, lower colon, urinary bladder and liver were slightly more highly exposed (all <2.3 mSv) than the other organs, and the breast dose was <0.3 mSv in every setup. Values of radiation exposure in 64- and 16-slice CTC differ only marginally when using the narrow collimation. In 64-slice CTC, the use of narrow (64 x 0.6 mm) collimation shows slightly elevated dose values compared to wider (20 x 1.2 mm) collimation.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT