Determination of platinum in human subcellular microsamples by inductively coupled plasma mass spectrometry
Erik Björn1, Yvonne Nygren, Tam Trinh Thi Nhu Nguyen
1Department of Chemistry, Umeå University, S-901 87 Umeå, Sweden. erik.bjorn@chem.umu.se
Abstract:
A fast and robust method for the determination of platinum in human subcellular microsamples by inductively coupled plasma mass spectrometry was developed, characterized, and validated. Samples of isolated DNA and exosome fractions from human ovarian (2008) and melanoma (T289) cancer cell lines were used. To keep the sample consumption to approximately 10 microl and obtain a high robustness of the system, a flow injection sample introduction system with a 4.6-microl sample loop was used in combination with a conventional pneumatic nebulizer and a spray chamber. The system was optimized with respect to signal/noise ratio using a multivariate experimental design. The system proved to be well suited for routine analysis of large sample series, and several hundreds of samples could be analyzed without maintenance or downtime. The detection limit of the method was 0.12 pg (26 pg/g) platinum. To avoid systematic errors from nonspectral interferences, it was necessary to use reagent matched calibration standards or isotope dilution analysis. An uncertainty budget was constructed to estimate the total expanded uncertainty of the method, giving a quantification limit of 2.3 pg (0.5 ng/g) platinum in DNA samples. The uncertainty was sufficiently low to study quantitative differences in the formation of Pt-DNA adducts after treatment with cisplatin using different exposure times and concentrations.
More Related Videos
14:53In Situ Detection and Single Cell Quantification of Metal Oxide Nanoparticles Using Nuclear Microprobe Analysis
Published on: February 3, 2018
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab