Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not
Gerard R Manecke1, William R Auger
1Department of Anesthesiology, UCSD Medical Center, San Diego, CA 92103, USA. gmanecke@ucsd.edu
Objective:
The purpose of this study was to evaluate the accuracy and precision of a novel algorithm that evaluates cardiac output by using arterial pressure waveform characteristics.
Design:
Prospective, observational study comparing the cardiac output values of intermittent thermodilution, continuous thermodilution, and continuous arterial pressure wave assessment.
Setting:
The intensive care unit in a tertiary care university hospital.
Participants:
Fifty postoperative cardiac surgical patients, within the first 12 hours after surgery.
Interventions:
All patients received a pulmonary artery catheter (PAC) and at least 1 systemic arterial pressure catheter. The data from the arterial catheter were processed by using a new arterial pressure cardiac output (APCO) algorithm. The data from the PAC (continuous and intermittent assessments) were collected for comparison with the APCO.
Measurements:
Two hundred ninety-five cardiac output measurements using intermittent thermodilution (ICO), continuous thermodilution (CCO), and arterial pressure-based output (APCO) were obtained during various times during the first 12 postoperative hours. The measurements of each method at each time point were compared by using Bland-Altman analysis.
Results:
The mean cardiac output ranged from 2.77 to 9.60 L/min. APCO, compared with ICO, revealed a bias of 0.55 L/min and precision of 0.98 L/min. APCO, compared with CCO, revealed a bias of 0.06 L/min and precision of 1.06 L/min. The APCO agreement between femoral and radial arterial catheters was close; the bias was -0.15 L/min, and the precision was 0.56 L/min.
Conclusions:
This novel arterial pressure cardiac output algorithm provides cardiac output assessments that agree satisfactorily for clinical purposes with intermittent and continuous thermodilution techniques in postoperative cardiac surgical patients. Further study is required for other patient populations and clinical situations.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessment of blood pressure in brachial artery(two-step method)