[Verification and application of the voxel-based Monte Carlo (VMC++) electron dose module of Oncentra MasterPlan]
Christian Scherf1, Josef Scherer, Ludwig Bogner
1Klinik und Poliklinik für Strahlentherapie, Klinikum der Johann Wolfgang Goethe-Universität, Theodor-Stern-Kai 7, 60590, Frankfurt/Main, Deutschland. christian.scherf@kgu.de
Unlabelled:
PURPOSE, MATERIAL AND METHODS: The dose calculation accuracy of the voxel-based Monte Carlo (VMC++) electron dose module of Oncentra MasterPlan (Nucletron B.V., Veenendaal, The Netherlands) was verified by measurements in homogeneous water phantoms.
Results:
Measured and calculated dose maxima on the central beam axis (calculations with 10,000-20,000 incident electron histories per cm(2)) agree well using standard applicator configurations as well as individually shaped inserts. Profile scans with higher electron energies (>/= 15 MeV) reveal differences up to 5% especially in the penumbra region. Depth dose curves agree best in the vicinity of maximum depths. In the buildup region energy-dependent differences up to 5% in both directions could be observed. In the decay region of depth dose curves calculated doses were up to 10% higher than measured values.
Conclusion:
Good VMC++ accuracy combined with moderate computing times of 1-15 min per beam satisfy all clinical needs. VMC++ allows, for the first time, accurate routine dose evaluations of radiation therapy with electrons. Adequate positioning of the dose reference point is essential. Even small displacements may significantly influence the calculation of monitor units.
Related Concept Videos
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...