Related Experiment Video
Updated: Jul 17, 2026

Atomic Force Microscopy Cantilever-Based Nanoindentation: Mechanical Property Measurements at the Nanoscale in Air and Fluid
Published on: December 2, 2022
Accuracy of the spring constant of atomic force microscopy cantilevers by finite element method
Bo-Yi Chen1, Meng-Kao Yeh, Nyan-Hwa Tai
1Department of Power Mechanical Engineering and Department of Materials Science and Engineering, National Tsing Hua University, Hsinchu 30013 Taiwan, Republic of China.
Abstract:
Atomic force microscopy (AFM) probe with different functions can be used to measure the bonding force between atoms or molecules. In order to have accurate results, AFM cantilevers must be calibrated precisely before use. The AFM cantilever's spring constant is usually provided by the manufacturer, and it is calculated from simple equations or some other calibration methods. The spring constant may have some uncertainty, which may cause large errors in force measurement. In this paper, finite element analysis was used to obtain the deformation behavior of the AFM cantilever and to calculate its spring constant. The influence of prestress, ignored by other methods, is discussed in this paper. The variations of Young's modulus, Poisson's ratio, cantilever geometries, tilt angle, and the influence of image tip mass were evaluated to find their effects on the cantilever's characteristics. The results were compared with those obtained from other methods.
Related Concept Videos
Euler's Formula to Columns: Problem Solving
The system comprises two vertical rigid bars, AB and BC, of...
Atomic Force Microscopy
The AFM Probe
The probe is regarded as the heart of any AFM setup and comprises the...