Pre-sterilisation cleaning of re-usable instruments in general dental practice
1Infection Research Group, Glasgow Dental Hospital and School, Glasgow, UK.
Objective:
This study examined the policies, procedures, environment and equipment used for the cleaning of dental instruments in general dental practice.
Materials And Methods:
A total of 179 surgeries were surveyed. This was an observational based study in which the cleaning processes were viewed directly by a trained surveyor. Information relating to surgery policies and equipment was also collected by interview and viewing of records. Data were recorded onto a standardised data collection form prepared for automated reading.
Results:
The BDA advice sheet A12 was available in 79% of surgeries visited. The most common method for cleaning dental instruments was manual washing, with or without the use of an ultrasonic bath. Automated washer disinfectors were not used by any surgery visited. The manual wash process was poorly controlled, with 41% of practices using no cleaning agent other than water. Only 2% of surgeries used a detergent formulated for manual washing of instruments. When using ultrasonic baths, the interval that elapsed between changes of the ultrasonic bath cleaning solution ranged from two to 504 hours (median nine hours). Fifty-eight percent of surgeries claimed to have a dedicated area for instrument cleaning, of which 80% were within the patient treatment area. However, in 69% of surgeries the clean and dirty areas were not clearly defined. Virtually all cleaning of dental instruments was undertaken by dental nurses. Training for this was provided mainly by demonstration and observed practice of a colleague. There was little documentation associated with training. Whilst most staff wore gloves when undertaking manual cleaning, 51% of staff did not use eye protection, 57% did not use a mask and 7% used waterproof overalls.
Conclusions:
In many dental practices, the cleaning of re-usable dental instruments is undertaken using poorly controlled processes and procedures, which increase the risk of cross infection. Clear and unambiguous advice must be provided to the dental team, especially dental nurses, on appropriate equipment, chemicals and environment for cleaning dental instruments. This should be facilitated by appropriate training programmes and the implementation of quality assurance procedures at each stage of the cleaning process.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene
Hand washing...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Asepsis