Related Experiment Video
Updated: Jul 16, 2026

Quantification of Levator Ani Hiatus Enlargement by Magnetic Resonance Imaging in Males and Females with Pelvic Organ Prolapse
Published on: April 17, 2019
Direction sensitive sensor probe for the evaluation of voluntary and reflex pelvic floor contractions
Christos E Constantinou1, Sadao Omata
1Stanford University Medical School, Stanford, California 94305, USA. ceconst@stanford.edu
Aims:
The development of a vaginal probe for the evaluation of the dynamics of pelvic floor function is described. Fundamental criteria in the design of this probe involves the incorporation of a means of assessing whether the isotonic forces closing the vagina are equally distributed or whether they are greater in some directions than others. The aim of this study is to present the design of directionally sensitive multi-sensor probe, having circumferential spatial resolution, constructed to identify the distribution of anisotropic forces acting on the vagina following voluntary and reflex pelvic floor contractions.
Materials And Methods:
Probe system consists of four pairs of force/displacement sensors mounted on leaf springs enabling isotonic measurements of voluntary and reflex contractions. Assembly is retractable to 23 mm for insertion, and expandable to 60 mm for measurement. Simultaneous measurements were made of force and displacement with the sensors oriented in the anterior/posterior and left/right orientation of the vagina. Using this probe, measurements were carried out to identify the temporal and spatial characteristic response of the vaginal wall. Data were analyzed with respect to voluntary pelvic floor and cough-induced contractions of nine subjects having a mean age of 64 years.
Results:
A robust probe system was developed and measurements were successfully made. Initial results show that the maximum force and displacement occurs during reflex contractions in the anterior aspect of the vagina validating the anisotropic nature of the forces acting on the vaginal wall. The data also show that both the force and displacement produced by the cough-induced has a higher magnitude than voluntary pelvic floor contraction.
Conclusions:
A directional multi-sensor vaginal probe has been developed to evaluate the force and displacement produced during isotonic pelvic floor contractions. Analysis of the results provided new biomechanical data demonstrating the anisotropic nature of vaginal closure as a consequence of pelvic floor contractions.
Related Concept Videos
The Micturition Reflex
The process begins with bladder filling, where the bladder wall stretches as urine accumulates. This stretching activates the urine storage reflex, mediated by the sacral spinal segments and the pontine storage center. Efferent sympathetic impulses stimulate the detrusor muscle to relax and the internal urethral sphincter to contract, facilitating urine...
Muscles of the Pelvic Floor and Perineum
Perineal Layer
The perineum is a diamond-shaped area below the pelvic diaphragm, divided into an anterior urogenital triangle that contains the external genitals and a posterior anal triangle housing the anus. The urogenital...
Reflex Activity
A reflex exam is a diagnostic procedure performed by a healthcare professional to evaluate the functionality of a patient's...