Microvolt T-wave alternans during atrial and ventricular pacing
Alaa A Shalaby1, Andrew Voigt, Aiman El-Saed
1Veterans Affairs Pittsburgh Healthcare System, Pittsburgh, Pennsylvania, USA. Alaa.Shalaby@med.va.gov
Ventricular pacing (VP) enables more microvolt T-wave alternans (MTWA) assessments than atrial pacing (AP), showing good concordance. VP yields higher amplitudes and noise, increasing non-negative results for sudden cardiac death risk evaluation.
Area of Science:
- Cardiology
- Electrophysiology
- Biomedical Engineering
Background:
- Microvolt T-wave alternans (MTWA) assessment using spectral analysis (SA) requires stable heart rates.
- Ventricular pacing (VP) can achieve the necessary heart rate stability for MTWA-SA.
- Atrial pacing (AP) is an alternative method for heart rate control during MTWA-SA.
Purpose of the Study:
- To compare the feasibility and diagnostic performance of MTWA-SA using short-term VP versus AP.
- To evaluate the concordance, sensitivity, specificity, and predictive values of MTWA-VP compared to MTWA-AP.
- To analyze differences in MTWA amplitude and noise levels between VP and AP.
Main Methods:
- Invasive electrophysiologic study in 42 patients undergoing risk evaluation for sudden cardiac death.
- Comparison of MTWA-SA results obtained during VP and AP.
- Evaluation of concordance (kappa statistic), sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV).
- Analysis of maximum MTWA amplitude (MValt) and noise levels in both pacing modes.
Main Results:
- MTWA-SA was completed in 40 patients with VP versus 32 patients with AP.
- VP-MTWA-SA demonstrated a sensitivity of 93%, specificity of 71%, NPV of 92%, and PPV of 72%.
- Results were concordant in 80% of patients (kappa = 0.62, P < 0.001).
- Noise levels were significantly higher during VP (1.4 ± 0.8) compared to AP (1 ± 0.8, P < 0.01).
- A trend towards higher MTWA amplitude was observed during VP (10.7 ± 5.3) versus AP (7.8 ± 3.9, P = 0.058).
Conclusions:
- MTWA-SA is more frequently completed using VP than AP.
- There is overall concordance between MTWA-VP and MTWA-AP.
- VP results in higher MTWA amplitudes and noise levels, leading to a greater proportion of non-negative findings.
More Related Videos
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrophysiology of Normal Cardiac Rhythm