Patient position verification with oblique radiation beams
Nanna M Sijtsema1, Alfons C M van den Bergh, Fred R Burlage
1Department of Radiation Oncology, University Medical Center Groningen/University of Groningen, Groningen, The Netherlands. n.m.sijtsema@rt.umcg.nl
Summary
Head and neck cancer radiotherapy patients
Area of Science:
- Radiation oncology
- Medical imaging
- Cancer treatment
Background:
- Accurate patient positioning is crucial in radiotherapy for head and neck cancers.
- Current verification methods often involve additional anterior-posterior (AP) and lateral beams, increasing treatment time and complexity.
Purpose of the Study:
- To evaluate if patient positioning during radiotherapy for head and neck cancer can be accurately determined using portal images from oblique radiation beams.
- To assess the potential to eliminate the need for supplementary AP and lateral verification beams.
Main Methods:
- Patient position was determined using portal images from oblique radiation beams.
- These positions were compared to those determined from standard AP and lateral verification beams.
- Analysis included 751 portal images from 18 patients.
Main Results:
- Setup errors in ventrodorsal and craniocaudal directions were accurately determined from oblique beams, comparable to AP/lateral beams.
- An additional AP beam was required for equivalent accuracy in the lateral direction due to beam angles.
- Position verification was more accurate with AP and lateral beams for patients receiving both oblique gantry angles and isocentric table rotations.
Conclusions:
- Oblique radiation beams can be effectively used for position verification in head and neck cancer radiotherapy when only oblique gantry angles are employed.
- This approach may streamline radiotherapy procedures by potentially replacing conventional verification beams in specific techniques.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
The most common cardiovascular diagnostic test is an X-ray. It produces images of the heart, blood vessels, and adjacent structures.
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Radiological investigations are paramount in the diagnosis and management of various pulmonary diseases. Two essential investigations are the Pulmonary Angiogram and the Positron Emission Tomography (PET) Scan.
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...