An investigation of the decontamination of Siqveland matrix bands

C L Whitworth1, K Davies, N O A Palmer

  • 1Hoghton Street Dental Practice, 22 Hoghton Street, Southport, UK. cwblackadder@btinternet.com

British Dental Journal
|February 20, 2007
PubMed
Abstract

Insights

Thorough cleaning of Siqveland matrix bands is not currently possible. Dental practitioners should discard matrix bands after single use to prevent blood contamination. Retainers can be effectively cleaned using enzymatic detergents and instrument washers.

Area of Science:

  • Dental materials science
  • Infection control in dentistry
  • Medical device reprocessing

Background:

  • Siqveland matrix bands and retainers are used in dental procedures.
  • Effective cleaning and sterilization of dental instruments are crucial for infection control.
  • Previous studies have evaluated cleaning methods for various dental instruments.

Purpose of the Study:

  • To investigate the efficacy of different pre-sterilization cleaning methods for Siqveland matrix bands and retainers.
  • To compare the effectiveness of enzymatic agents, washer-disinfectors, and instrument washers in removing blood contamination.
  • To assess the possibility of cleaning assembled Siqveland matrix bands and retainers.

Main Methods:

  • Artificial and clinical contamination of Siqveland matrix bands and retainers with blood.
  • Application of enzymatic agents, washer-disinfectors, and instrument washers for cleaning.
  • Measurement of residual blood contamination using a modified Kastle-Meyer test.
  • Comparison of cleaning methods against an unprocessed control group.

Main Results:

  • Residual blood was detected on all clinically contaminated Siqveland matrix bands and retainers.
  • Washer-disinfector processing was most effective for artificially contaminated bands.
  • Enzymatic pre-soaking followed by instrument washing was most effective for clinically contaminated bands and retainers.
  • Significant blood volumes (0.13-7.1 microl for bands, 0.001-1.523 microl for retainers) remained after cleaning.

Conclusions:

  • Current cleaning methods are insufficient for assembled Siqveland matrix bands.
  • Siqveland matrix bands should be considered single-use items.
  • Siqveland retainers can be effectively cleaned after band removal using enzymatic pre-soaking and instrument washing.