Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and
M Francone1, A Napoli, I Carbone
1Dipartimento di Scienze Radiologiche, Università degli Studi di Roma La Sapienza, Policlinico Umberto I, Viale Regina Elena 324, I-00161 Rome, Italy, Marco.Francone@uniroma1.it
Purpose:
We present our initial clinical experience with a recently introduced 64-detector computed tomography (64-MDCT) scanner that makes use of a periodic motion of the focal spot in the longitudinal direction (z-flying focal spot), which enables it to reach a final spatial resolution of 0.4 x 0.4 x 0.4 mm(3) and a temporal resolution of 83 ms.
Materials And Methods:
A total of 114 patients (108 men, six women; age range 36-77 years, mean 63.1 years) underwent retrospective electrocardiogram (ECG)-gated examination of the coronary arteries using a 64-MDCT scanner (Somatom Sensation 64, Siemens Medical Solutions, Germany). Acquisition parameters were the following: collimation 64 x 0.6 mm, 800 quality reference milliampere second (mAs), 120 kVp, 0.33-s gantry rotation time and pitch 0.2. Images were acquired in all cases after i.v. administration of 80 ml of contrast agent (Iomeron 400 mgI/dl, Bracco, Italy) + 30 ml of saline at 4 /s and delay time determined using a bolus triggering technique. Oral betablockers were administered to patients with heart rate (HR) >75 bpm. To reduce radiation exposure, an automatic exposure control system was applied in all cases to adapt tube current to patient size and anatomic shape (CARE Dose 4D, Siemens Medical Solutions, Germany). The optimal temporal window for raw data reconstruction was chosen from an initial preview of images reconstructed with different phase settings (range 0%-95% RR interval with 5% gap) at a selected anatomical level in the mid part of the right coronary artery. CT dose index volume and effective dose were quantified in all patients using dedicated software.
Results:
Mean HR recorded during image acquisition was 65.6+/-19.2 bmp (range: 44-96 bmp), and beta-blockers were administered to 16/114 patients (14.0%). Technical adequacy was achieved in all patients but two (2/114; 1.7%). In patients with HR <60 bmp, the best reconstruction intervals were identified in the end-systolic (30%-35% of the RR interval) and end-diastolic (60%-65% of the RR interval) phases; with faster HR (>80 bmp), high image quality was observed in end-systole (30%-35% of the RR interval). Mean CT dose index (CTDI) volume was 36.53+/-8.30 mGy per patient. In comparison with a conventional examination with fixed mAs, the use of the CARE Dose 4D system provided a 33.3% CTDI volume reduction (p<0.001). Mean effective dose was 9.5+/-3.4 millisievert (mSv) per patient (range 7.1-17.7).
Conclusions:
The 64-MDCT scanner diagnostic performance for coronary CT angiography is further improved with better spatial and temporal resolution and faster scan times; besides, initial clinical results are promising. The use of dose-reducing acquisition techniques is mandatory to limit radiation exposure to the patient.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI