Radiographic failure patterns of polished cemented stems.
Brian A Jewett1, Dennis K Collis
1Orthopedic Healthcare Northwest, 1200 Hilyard Street, Suite 600, Eugene, OR 97401, USA. Brian.Jewett@OHN.com
Clinical Orthopaedics and Related Research
|February 22, 2007
Summary
Polished femoral stems in total hip arthroplasty showed low aseptic loosening rates, regardless of geometry. Design changes did not significantly alter failure patterns or improve outcomes compared to earlier rough stems.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Materials science in implants
Background:
- Femoral component design in total hip arthroplasty (THA) has evolved to enhance implant longevity.
- Surface finish and geometry are key factors influencing cemented femoral stem performance.
- Previous studies indicated potential issues with rough femoral stems.
Purpose of the Study:
- To evaluate the failure rates and radiographic failure patterns of four distinct polished femoral stem geometries.
- To determine if design modifications in polished stems over a decade improved clinical outcomes.
- To compare the performance of polished stems to historical data on rough stems.
Main Methods:
- Retrospective review of 1031 THAs performed between 1993 and 2003.
- Analysis of aseptic loosening rates and failure patterns in polished femoral stems with 4-microinch surface roughness but varying geometries.
- Minimum 2-year follow-up for all patients.
Main Results:
- Seven stems (0.6%) required revision for aseptic loosening by 2005, all originating at the cement-bone interface.
- Average time from loosening onset to revision was 3.3 years.
- No significant differences in loosening rates or radiographic failure patterns were observed among the four stem geometries. Bone damage was minimal and manageable.
Conclusions:
- Polished femoral stems, irrespective of geometric variations, demonstrate a low incidence of aseptic loosening in THA.
- The design evolution of polished stems did not yield a discernible improvement in failure rates or patterns compared to earlier designs.
- These findings suggest that polished femoral stems offer a favorable outcome with less bone damage than previously reported for rough stems.