Adverse impact of temperature and humidity on blood glucose monitoring reliability: a pilot study
Michael J Haller1, Jonathan J Shuster, Desmond Schatz
1Division of Pediatric Endocrinology, University of Florida, Gainesville, Florida 32610, USA. hallemj@peds.ufl.edu
Background:
The effect of temperature and humidity on the reliability of glucose monitoring systems (GMS) has not been well reported. We performed a study to determine if the effect of temperature and humidity on glucose values could be accurately predicted.
Methods:
Eight GMS, including meter, strips, and glucose control solution (GCS), were placed in an environmental chamber for 50 days. Failure of the environmental chamber allowed for the serendipitous observation of the effect on glucose results of a range of temperature (54-87 degrees F) and humidity (49-100%) typically experienced by patients when performing self-monitoring of blood glucose. GCS was used to generate glucose results in triplicate for each meter daily. Temperature and humidity variations were recorded, and a linear model was generated to explain the variation in glucose results.
Results:
Both temperature and humidity had significant effects on the reliability of nearly all GMS. The effect of temperature was greater than the effect of humidity.
Conclusions:
In this pilot study, GMS were unreliable at temperatures and humidity within the manufacturers' recommended limits. Such variations could result in clinically significant errors in diabetes management. Further studies are needed to confirm the potential inadequacies of GMS with multiple meters and real-time patient blood samples. Glucose meter manufacturers should develop systems that either are less affected by climate or adjust reported values based on ambient temperature and humidity.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Hyperglycemia
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...