High dose-rate brachytherapy source position quality assurance using radiochromic film
M D C Evans1, S Devic, E B Podgorsak
1Medical Physics Department, McGill University Health Centre, Montreal, Quebec, Canada. mevans@medphys.mcgill.ca
Abstract:
Traditionally, radiographic film has been used to verify high-dose-rate brachytherapy source position accuracy by co-registering autoradiographic and diagnostic images of the associated applicator. Filmless PACS-based clinics that do not have access to radiographic film and wet developers may have trouble performing this quality assurance test in a simple and practical manner. We describe an alternative method for quality assurance using radiochromic-type film. In addition to being easy and practical to use, radiochromic film has some advantages in comparison with traditional radiographic film when used for HDR brachytherapy quality assurance.