Related Experiment Video
Updated: Jul 16, 2026

Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Measurement of pressure-time product during spontaneous assisted breathing by rapid interrupter technique
Giacomo Bellani1, Nicolò Patroniti, Dieter Weismann
1Dipartimento di Medicina Sperimentale, Ambientale e Biotecnologie Mediche, Università degli Studi Milano Bicocca, Ospedale San Gerardo, Italy.
Background:
Measuring the work of breathing of patients undergoing spontaneous assisted ventilation can be useful to monitor and titrate ventilatory support. The aim of this study was to obtain measurements of the pressure generated by the respiratory muscles (PMUSC) and the derived pressure-time product (PTP; a good indicator of the metabolic work of breathing), performing the rapid interrupter technique with a commercial ventilator.
Methods:
A Draeger Evita 4 ventilator (Draeger Medical, Lubeck, Germany) was controlled by a personal computer to rapidly interrupt the airway flow at different times and volumes of the respiratory cycle during pressure-support ventilation. From the airway pressure tracing after the occlusion, the authors estimated the alveolar pressure and PMUSC; the integration of PMUSC values over the inspiratory time yields the measurement of PTP. Esophageal pressure measurements were used as a reference. After a bench study of the valves' performance, the authors performed 11 measurement sequences in eight patients.
Results:
The closure times for the inspiratory and expiratory valves were 74 +/- 10 and 61 +/- 13 ms, respectively. The interrupter technique provided a reliable estimate of PMUSC (PMUSC, occl = 1.00 . PMUSC, pes + 0.19; r = 0.88; 95% confidence interval for agreement, +5.49/-5.32 cm H2O). PTPoccl tightly correlated with PTPpes (PTPoccl = 0.95 . PTPpes + 0.13; r = 0.96; 95% confidence interval, 1.94/-1.61 cm H2O . s).
Conclusion:
The rapid interrupter technique can be performed by means of a commercial ventilator, providing reliable measurement of PMUSC and PTP.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Measurement of Blood Pressure
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)