Reexamining the quantification of perfusion MRI data in the presence of bolus dispersion
Linda Ko1, Marina Salluzzi, Richard Frayne
1Department of Electrical and Computer Engineering, University of Calgary, Calgary, Alberta, Canada.
Purpose:
To determine the true impact of dispersion upon cerebral blood flow (CBF) quantification by removing an algorithm implementation-induced systematic error.
Materials And Methods:
The impact of dispersion on the arterial input function (AIF) between measurement and entry into the tissue of interest on CBF estimates was simulated assuming: 1) contralateral circulation flow that introduces a true arterial tissue delay (ATD)-related dispersive component; and 2) the presence of an arterial stenosis that disperses and shifts the AIF peak entering the tissue; increasing the apparent ATD relative to the original AIF.
Results:
Previously reported CBF estimates for the stenosis dispersion model were found to be a mixture of true dispersive effects and an algorithm implementation-induced systematic error. The true CBF(MEASURED)/CBF(NO-DISPERSION) ratios for short mean transit times (MTT) (normal) and long MTT (infarcted) tissue were similar for both dispersion models evaluated; this was an unanticipated result. The CBF quantification inaccuracies induced through the dispersion model truly related to ATD were lower than for the local stenosis-based dispersion for small ATD values.
Conclusion:
Correcting the systematic error present in a previous deconvolution study removes the reported ATD-related impact on CBF quantification. The impact of dispersion was smaller than half that reported in previous simulation studies.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Magnetic Resonance Imaging
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...