Corrective measures for compromised oxygen delivery during endotracheal tube cuff deflation with high-frequency
Patrick F Allan1, Gregory Naworol
1Department of Respiratory Therapy, Wilford Hall Medical Center, 759th MCCP, 2200 Bergquist Drive, Lackland Air Force Base, TX 78236, USA. patrick.allan@lackland.af.mil
Objective:
To determine the effect of endotracheal-tube cuff deflation on airflow and F(IO2) during high-frequency percussive ventilation (HFPV), and explore methods of correcting the cuff-deflation-associated decrease in mean airway pressure and F(IO2) at the carina.
Methods:
Using a mechanical lung model in our respiratory research laboratory, we measured circuit pressure near the connection to the endotracheal tube (P(vent)), mean airway pressure (P(aw)), pulsatile tidal volume (V(T)), and F(IO2) at the artificial carina. During cuff deflation we manipulated the pulsatile frequency, pulsatile flow, and the HFPV integral nebulizer. We then assessed 4 methods of correcting the decreased F(IO2) and airway pressure during cuff deflation: (1) oxygen delivery at the inspiratory fail-safe valve, (2) oxygen delivery at the T-piece between the HFPV and the endotracheal tube, (3) continuous activation of the HFPV's integral nebulizer, and (4) oxygen insufflation into the suction channel of the endotracheal tube.
Results:
Cuff deflation reduced P(vent), P(aw), pulsatile V(T), and F(IO2). Increasing the pulsatile flow and decreasing the pulsatile frequency further reduced F(IO2) during cuff deflation. Injecting supplemental oxygen at the inspiratory fail-safe valve provided the best F(IO2) increase. Injecting oxygen at the T-piece provided the second best F(IO2) increase. Continuous activation of the integral nebulizer provided the third best F(IO2) increase. Oxygen insufflation to the suction channel was least effective in correcting the F(IO2) decrease caused by cuff deflation.
Conclusion:
Cuff-deflation-associated F(IO2), P(aw), and pulsatile V(T) compromise can be partially corrected by any of the 4 methods we studied. Injecting supplemental oxygen at the inspiratory fail-safe valve is the most effective method.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...