Assessing data quality in manual entry of ventilator settings
David K Vawdrey1, Reed M Gardner, R Scott Evans
1Department of Biomedical Informatics, University of Utah School of Medicine, Salt Lake City, UT 84112-5750, USA. david.vawdrey@hsc.utah.edu
Objective:
To evaluate the data quality of ventilator settings recorded by respiratory therapists using a computer charting application and assess the impact of incorrect data on computerized ventilator management protocols. DESIGN An analysis of 29,054 charting events gathered over 12 months from 678 ventilated patients (1,736 ventilator days) in four intensive care units at a tertiary care hospital.
Measurements:
Ten ventilator settings were examined, including fraction of inspired oxygen (Fio (2)), positive end-expiratory pressure (PEEP), tidal volume, respiratory rate, peak inspiratory flow, and pressure support. Respiratory therapists entered values for each setting approximately every two hours using a computer charting application. Manually entered values were compared with data acquired automatically from ventilators using an implementation of the ISO/IEEE 11073 Medical Information Bus (MIB). Data quality was assessed by measuring the percentage of time that the two sources matched. Charting delay, defined as the interval between data observation and data entry, also was measured.
Results:
The percentage of time that settings matched ranged from 99.0% (PEEP) to 75.9% (low tidal volume alarm setting). The average charting delay for each charting event was 6.1 minutes, including an average of 1.8 minutes spent entering data in the charting application. In 559 (3.9%) of 14,263 suggestions generated by computerized ventilator management protocols, one or more manually charted setting values did not match the MIB data.
Conclusion:
Even at institutions where manual charting of ventilator settings is performed well, automatic data collection can eliminate delays, improve charting efficiency, and reduce errors caused by incorrect data.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.