Related Experiment Video
Updated: Jul 16, 2026

Fluorescence and Bioluminescence Imaging of Subcellular Ca2+ in Aged Hippocampal Neurons
Published on: December 1, 2015
Fluorescence absorbance inner-filter decomposition: the role of emission shape on estimates of free Ca(2+) using
Paul R Territo1, Jeremy Heil, Salil Bose
1Lilly Center for Molecular and Anatomical Imaging, Lilly Research Laboratories, Eli Lilly and Company, 2001 West Main Street, B220/GL54, Greenfield, Indiana 46140, USA. territop@lilly.com
Abstract:
A method for decomposing complex emission spectra by correcting for known inner-filter effects is described. This approach builds on previous work using a linear combination of model emission spectra and combines the known absorption characteristics of the system to fit the composite emission spectrum. Rhod-2, which has a small Stokes shift and significant self-absorption, was used as the model system. By adding the absorption characteristics of Rhod-2 to the model, the degree of fit was significantly improved, thus minimizing residuals, and accurately predicted the spectral shape changes with increasing concentration, [Rhod-2]. More complex studies were conducted with Rhod-2 in isolated cardiac mitochondria with multiple emission and absorption elements. By including known absorbances to the spectral decomposition, the overall precision increased almost four fold. Moreover, this approach eliminated the significant [Rhod-2] dependence on the apparent K(50) and therefore improved the accuracy of free [Ca(2+)] calculations. These data demonstrate that secondary inner-filter correction can significantly improve spectral decomposition of complex emission spectra, which are used in a variety of biological applications.