Objective:
Future automated external defibrillators are being designed to direct rescue efforts (chest compressions first vs. defibrillation) by inferring the duration of ventricular fibrillation based on its waveform characteristics such as frequency content. This approach assumes that the ventricular fibrillation waveform is an appropriate surrogate for ventricular fibrillation duration and is not affected by structural heart disease. We hypothesized that an acute myocardial infarction may alter the frequency content of ventricular fibrillation.
Design:
Animal intervention study with comparison to control group.
Setting:
University animal laboratory.
Subjects:
Twenty-seven swine.
Interventions:
Acute myocardial infarction was induced by occlusion of the mid-left anterior descending artery. Ventricular fibrillation was induced in swine with acute myocardial infarction and control swine.
Measurements And Main Results:
Ventricular fibrillation was induced in 11 swine with an acute myocardial infarction and in 16 control swine. Ventricular fibrillation waveforms were analyzed for mean, median, and dominant frequency, as well as bandwidth and amplitude. All frequency characteristics were significantly (p < .001) altered in swine with acute myocardial infarction compared with controls. Specifically, these characteristics were significantly depressed and varied little over time in swine with acute myocardial infarction compared with controls.
Conclusions:
These data establish that ventricular fibrillation during an acute myocardial infarction has an altered frequency content and time evolution compared with ventricular fibrillation without coronary obstruction. Frequency characteristics such as mean, median, dominant, and bandwidth show little variation in time after an acute myocardial infarction and are not suitable surrogates for ventricular fibrillation duration. These findings have important implications for the development of "smart" automated external defibrillators designed to determine duration of ventricular fibrillation from the waveform characteristics.
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Mechanism of Cardiac Arrhythmias