Related Experiment Video
Updated: Aug 16, 2026

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
One-step, labeled-antibody assay for measuring free thyroxin. II. Performance in a multicenter trial
C P Sheehan1, N D Christofides
1Amerlite Diagnostics Ltd., Cardiff Laboratories, Whitchurch, U.K.
Abstract:
The technical and diagnostic performance of a labeled antibody radioassay for free thyroxin (Amerlex-MAB* FT4) was examined in a multicenter trial. The overall mean within-assay precision (CV) in eight centers was 6.2% at 30 pmol/L and 3.6% at 13 pmol/L concentrations of FT4. Between-assay precision was 5.4% at 18 pmol/L and 6.8% at 50 pmol/L. The euthyroid central 95% reference range for FT4 was 11.3 to 24.3 pmol/L. Results of the method correlated well with those of an analog radioimmunoassay: [Amerlex-MAB* FT4] = 1.09 [Amerlex-M FT4] + 1.50 pmol/L (r = 0.96, n = 732). Clinical performance of the assay was better than that of first-generation analog tracer assays for sera from patients with nonthyroidal illness or binding-protein abnormalities.