Related Experiment Video
Updated: Jul 16, 2026

08:01
Detection of Invasive Pulmonary Aspergillosis in Haematological Malignancy Patients by using Lateral-flow Technology
Published on: March 22, 2012
Histoplasmosis-associated cross-reactivity in the BioRad Platelia Aspergillus enzyme immunoassay
L Joseph Wheat1, Emily Hackett, Michelle Durkin
1MiraVista Diagnostics, 4444 Decatur Blvd., Suite 300, Indianapolis, IN 46241, USA. jwheat@miravistalabs.com
Clinical and Vaccine Immunology : CVI
|March 9, 2007
Abstract:
We observed false-positive results in the Platelia Aspergillus enzyme-linked immunoassay (EIA) for specimens from patients with histoplasmosis and mice with experimental infection. Platelia Aspergillus EIA-positive specimens were negative in the second-generation Histoplasma antigen EIA. Care must be taken to exclude histoplasmosis for patients with positive Platelia Aspergillus EIA results.