Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Technical optimization of four-channel multidetector-row helical computed tomography for depicting arterial stenosis:
S Takahashi1, M Takamura, S Yamamoto
1Department of Clinical Informative Imageology, Osaka University Graduate School of Medicine, Osaka, Japan. s.takahashi@rad.umcn.nl
Purpose:
To determine the effects of detector configuration, as well as vessel orientation, on the depiction accuracy of arterial stenosis using four-channel multidetector-row helical computed tomography (MDCT) angiography in vitro.
Material And Methods:
Five acrylic vessel phantoms (3 mm in diameter with 25 or 50% stenosis, or 5 mm with 25, 50, or 75% stenosis) were scanned with a four-channel MDCT scanner at five vessel orientations (0, 30, 45, 60, and 90 degrees to the z-axis) using 4 x 1.25, 2.5, 3.75, and 5.0-mm detector configurations at beam pitches of 0.75 and 1.5. The percentage of stenosis was calculated by the ratio of the full width at half maximum for stenotic and non-stenotic portions of the phantom, and compared to the actual known values.
Results:
A detector configuration of 4 x 1.25 mm provided good reproducibility, as well as high accuracy for assessing vessel stenosis, while a 4 x 2.5-mm or wider detector configuration caused underestimations of stenosis. Although the phantoms perpendicular to the z-axis were underestimated, the errors were kept in clinically acceptable ranges using the 4 x 1.25-mm detector configuration.
Conclusion:
Four-channel MDCT accurately discerns stenosis for vessel phantoms of 3 or 5 mm in diameter at any orientation when using a detector configuration of 4 x 1.25 mm.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography