Related Experiment Video
Updated: Jul 16, 2026

Performing Spectroscopy on Plasmonic Nanoparticles with Transmission-Based Nomarski-Type Differential Interference Contrast Microscopy
Published on: June 5, 2019
Analysis of total uncertainty in spectral peak measurements for plasmonic nanoparticle-based biosensors
Adam Curry1, Gregory Nusz, Ashutosh Chilkoti
1Department of Biomedical Engineering, Duke University, Durham, North Carolina 27708, USA. accurry@duke.edu
Abstract:
One goal of recent research on plasmonic nanoparticle-based sensors is maximizing nanoparticle sensitivity or shift of resonance peak wavelength per refractive index change. Equally important is a measurement system's peak location uncertainty or shift resolution. We provide systematic analyses and discuss optimization of factors that determine peak location uncertainty, reporting values as low as 0.3 nm for the presented scheme. This type of analysis is important, in part, because it provides a means of evaluating detection thresholds for biosensor applications such as analyte binding. We estimate thresholds of 310 streptavidin molecules for the presented scheme and 20 molecules with system improvements.