Related Experiment Video
Updated: Jul 16, 2026

Laparoscopic Extracorporeal Knot-Tying for Uterine Vessel Occlusion during Hysterectomy with Cervical Cerclage in Large Uteri
Published on: September 12, 2025
Late bleeding problems with the levonorgestrel-releasing intrauterine system: evaluation of the endometrial cavity
Maria Rönnerdag1, Viveka Odlind
1Department of Women's and Children's Health, University of Uppsala, S 75267 Uppsala, Sweden. Maria.ronnerdag@comhem.se
Background:
This study was conducted to determine causes of abnormal bleeding patterns occurring after at least 6 months of successful use of the levonorgestrel-releasing intrauterine system (LNG-IUS).
Study Design:
Matched-pair, case-control study. Fifteen users of the device with unacceptable bleeding problems (cases) and 15 controls using the device but without such problems were recruited. All women had vaginal ultrasonography and hysteroscopy performed to evaluate the endometrial cavity.
Results:
An abnormality demonstrated by ultrasonography or hysteroscopy (either device displacement or leiomyomas) was more common among cases than controls. The McNemar's test for discordant case-control pairs yielded a p value of .046. After removal of the device and reinsertion of another levonorgestrel-releasing device, all cases reported improvement of the bleeding pattern.
Conclusions:
Displacement of the device and intracavitary leiomyomas are commonly associated with unacceptable bleeding in women after prolonged use of the LNG-IUS. Ultrasonography and hysteroscopy may be useful in evaluating bleeding complaints in long-term users of this device.
Related Concept Videos
Intrauterine Drug Delivery Systems
Menses Phase
When fertilization does not occur, the corpus luteum deteriorates, causing a significant drop in the levels of estrogen and progesterone in the body. This hormonal decrease triggers the release of prostaglandins, which cause the uterine...
Ovarian Cycle
Birth Control Methods