Tape-stripping method in man: comparison of evaporimetric methods
Hongbo Zhai1, Emi Dika, Marina Goldovsky
1Department of Dermatology, University of California, School of Medicine, San Francisco, CA 94143-0989, USA. zhaih@derm.ucsf.edu
Both open and closed chamber devices provide similar transepidermal water loss (TEWL) measurements on stripped human skin. This study confirms their comparable performance in a tape-stripping model, offering reliable data for skin barrier research.
Area of Science:
- Dermatology
- Biophysics
Background:
- Closed chamber evaporimeters risk saturation with prolonged occlusion.
- Previous comparisons of open and closed chamber devices were on intact skin.
- Data on stripped skin with higher evaporation rates are limited.
Purpose of the Study:
- To compare the sensitivity and correlation of open and closed chamber devices.
- To evaluate TEWL measurements in a human tape-stripping model.
- To quantify stratum corneum (SC) removed during tape stripping.
Main Methods:
- Ten healthy volunteers participated in a randomized, paired comparison study.
- Transepidermal water loss (TEWL) was measured using both open and closed chamber devices.
- Twenty tape strippings were performed on each forearm, with measurements taken after 10 and 20 strippings.
Main Results:
- Mean TEWL values were similar between the open and closed chamber devices after 10 and 20 tape strippings.
- No statistically significant differences were observed in TEWL measurements between the two devices.
- Stratum corneum (SC) aggregates in tape strips were comparable and showed no significant differences.
Conclusions:
- Both open and closed chamber evaporimeters are suitable for measuring TEWL on stripped human skin.
- The devices demonstrate comparable performance in a tape-stripping model.
- Findings support the use of either device for in vivo studies of skin barrier function.
Related Concept Videos
Volatilization
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Titrimetric Methods: Types and Commonly Used Strategies
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Distillation: Vapor–Liquid Equilibria
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...