The validity of bispectral index values from a dislocated sensor: a comparison with values from a sensor located in
Toshinori Horiuchi1, Masahiko Kawaguchi, Naoko Kurita
1Department of Anesthesiology, Nara Medical University, Nara, Japan.
Background:
The influence of sensor dislocation on bispectral index (BIS) values is not clear. We compared the BIS values obtained from dislocated sensors with those from the commercially recommended positions.
Methods:
We used two BIS sensors for each patient receiving propofol-based anesthesia; one in the recommended position and one positioned around the lateral corner of the right eye.
Results:
Bland and Altman analysis revealed better agreement of two BIS values when the values during induction of and emergence from anesthesia were excluded.
Conclusions:
The results indicate that during induction of and emergence from general anesthesia, a dislocated BIS sensor may produce questionable BIS values.
More Related Videos
04:05Clinical Application of Phase Angle and BIVA Z-Score Analyses in Patients Admitted to an Emergency Department with Acute Heart Failure
Published on: June 30, 2023
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...