Related Experiment Video
Updated: Jul 16, 2026

Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine
Published on: January 25, 2019
Evaluation of the high-sensitivity, full-range Olympus CRP OSR6199 application on the Olympus AU640
Anne Marie Dupuy1, Adeline Boutet, Jean Paul Cristol
1Department of Biochemistry, Lapeyronie Hospital, Montpellier, France.
Background:
Implementation of high-sensitivity C-reactive protein (hs-CRP) assays as a routine laboratory parameter may be necessary. A single CRP method giving reliable results for the whole concentration range (0.1-160 mg/L) is the most practical solution for a laboratory. The aim of this study was to assess the Olympus CRP full-range immunoturbidimetric assay on an Olympus AU640 biochemistry analyser.
Methods:
CRP was determined in a population of patients simultaneously with the Randox CRP immunoturbidimetric assay and two immunonephelometric assays, the Dade Behring and Beckman Coulter CRP, on BN 100 and Immage systems, respectively.
Results:
Analytical performance, including precision and linearity, was excellent. The correlation coefficient for linear regression was r(2)=0.998, indicating that the assay method exhibited good linearity in the working range 0-160 mg/L. The intra- and inter-assay CVs for the Olympus CRP method were <3.6% over a wide range of CRP concentrations. Linear regression analysis indicated excellent agreement between CRP values for all patients, with correlation coefficients of r(2)>0.996, 0.993 and 0.989 for the Randox, Dade Behring and Beckman methods, respectively. Bland-Altman analyses demonstrated the excellent concordance of the CRP methods for patient samples.
Conclusions:
This study demonstrates that the new full-range Olympus method can be applied to measure CRP as a marker of inflammation and of cardiovascular risk.