Fizzy Extraction of Volatile Organic Compounds Combined with Atmospheric Pressure Chemical Ionization Quadrupole Mass Spectrometry
Published on: July 14, 2017
The effect of analyte acidity on signal suppression and the implications to peak purity determinations using
Kristine Liao1, Brian W Pack, Nicholas P Toltl
1Eli Lilly Canada Inc., Toronto, Ontario, Canada M1N 2E8.
Abstract:
The effect of a co-eluting halogenated phenol, spiked at 1% of the main analyte level, has been examined for a series of halogenated phenols using LC-MS techniques. Similarly, the effect of co-eluting anilines has been investigated. The purpose of the work presented here was to evaluate the degree of signal suppression for structurally similar halogenated phenols and for similar anilines utilizing atmospheric pressure chemical ionization (APCI) in the negative mode and electrospray (ESI) in positive mode, respectively. A correlation between the effects of analyte ionization efficiency resulting from co-eluting compounds (signal suppression) and pK(a) has been made for these compounds. It was found that minimal signal suppression occurs when the spiked impurity has a similar (Delta pK(a)<1.5) acidity when compared to the main peak it is co-eluting with. The degree of signal suppression sharply increases when the difference in pK(a)'s between the main peak and the spiked impurity was greater than 1.5 units. Thus, when the main peak is much less acidic (more than 1.5 pK(a) difference) than the co-eluting impurity, signal suppression of the latter would not occur in negative mode APCI. Similarly, when the main peak is much less basic than the co-eluting peak, signal suppression of the impurity will also not be found for aniline compounds in positive mode ESI. Furthermore, the degree of signal suppression decreases as a function of sample load such that injections of 3 microg or less show no discernible impact on the spiked impurity peak. Ultimately, these results indicate that the use of mass spectrometry (MS) in peak purity determinations requires numerous considerations prior to assessing main peak purity. The optimization of sample load during an impurities assay will maximize co-eluting impurity signal as purity determinations by mass spectrometry made at sample loads above the 3 microg (sample load) threshold increase the risk for false negative assessment of impurities.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Chemical Ionization (CI) Mass Spectrometry
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Overview
Atomic Emission Spectroscopy: Interference