Background:
Pacemaker (PM) system implantation in neonates and infants is often complicated by hemodynamic instability, small vessel size, and abnormal cardiovascular anatomy. Thus, an open surgical approach for epicardial lead insertion is often required. We assessed the long-term outcomes after epicardial PM implantation in this age group.
Methods:
Between 1992 and 2004, 22 consecutive patients underwent PM implantation within the first year of life. Bipolar steroid-eluting epicardial leads (Medtronic CapSure Epi 10366 and 4968) were inserted through median sternotomy, the sybxyphoid approach, or thoracotomy, and connected to various pulse generators.
Results:
Pacemakers were implanted at a median age of 35 days (range, 1 to 300). Intracardiac anatomy was abnormal in 17 patients. Indications for PM therapy were heart block in 18 patients and sinus node dysfunction in 4 patients. During a median follow-up of 4.6 years (range, 4 days to 12.8 years), 7 devices were replaced owing to end of battery life (n = 6) or elective device repositioning (n = 1), at a median of 4.1 years (range, 1 to 7.8). One dislodged ventricular lead and 2 atrial lead sensing failures were observed. Sensing, pacing thresholds, and lead impedances showed good implant and stable follow-up values.
Conclusions:
Pacemaker-associated morbidity is low. Pacemaker system complications with epicardial leads are rare. Battery life is relatively shorter compared with children and adults because of the fast heart rate and complete PM dependency in most of these children. Even for neonates and infants, modern pacemaker therapy is feasible, safe, and effective.
More Related Videos
06:29Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiomyopathy V: Interprofessional Care