Related Experiment Video
Updated: Jul 16, 2026

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
Development of the technical capabilities needed to build and position a prepolarization coil for a magnetic
Abstract:
An experiment to show that a magnetic resonance imaging (MRI) magnet could be assembled around a patient, and used as part of a prepolarization system in which substantial transient forces are applied to parts of it, is described. The paper describes the circumstances that develop as a result of the application of the large transient fields used in this type of study, and outlines the reason for the tolerances that are permissible on the alignment of the system components. It then describes a test rig used to evaluate how the various problems might be overcome, and reports on the performance achieved with this rig. On the basis of this work, it appears that a system could be developed that would allow the application of these methods in clinical MRI.
Related Concept Videos
Magnetic Resonance Imaging
Magnetism
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
Atomic Nuclei: Magnetic Resonance
Magnetic Field Of A Current Loop
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Atomic Nuclei: Nuclear Relaxation Processes