Related Experiment Video
Updated: Jul 16, 2026

Determination of 45 Pesticides in Avocado Varieties by the QuEChERS Method and Gas Chromatography-Tandem Mass Spectrometry
Published on: December 8, 2023
Determination of synthetic acaricides residues in beeswax by high-performance liquid chromatography with photodiode
Sabine Adamczyk1, Regina Lázaro, Consuelo Pérez-Arquillué
1Department of Animal Production and Food Science, Veterinary Faculty, University of Zaragoza, Miguel Servet 177, 50013 Zaragoza, Spain.
Abstract:
A multiresidue HPLC method for identification and quantification of the synthetic acaricides fluvalinate, coumaphos, bromopropylate and its metabolite 4,4'-dibromobenzophenone in beeswax has been developed. Different techniques were tested and modified. The method consists of a sample preparation with isooctane followed by solid phase extraction using Florisil columns. Determination of the synthetic acaricides is achieved by HPLC with a photodiode array detector. Analytical performance of the proposed method, including sensitivity, accuracy and precision was satisfactory. The LOD for the analytes varied between 0.1 and 0.2 microg g(-1) wax and the recoveries between 70 and 110%. Relative standard deviation of the repeatability of the method is <15% and reproducibility is <31%.