Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Sensitive and ultra-fast determination of arsenic(III) by gas-diffusion flow injection analysis with
Cristina Lomonte1, Matthew Currell, Richard J S Morrison
1School of Chemistry, The University of Melbourne, Victoria 3010, Australia.
Abstract:
A novel chemiluminescence gas-diffusion flow injection system for the determination of arsenic(III) in aqueous samples is described. The analytical procedure involves injection of arsenic(III) samples and standards into a 0.3 mol L(-1) hydrochloric acid carrier stream which is merged with a reagent stream containing 0.2% (w/v) sodium borohydride and 0.015 mol L(-1) sodium hydroxide. Arsine, generated in the combined carrier/reagent donor stream, diffuses across the hydrophobic Teflon membrane of the gas-diffusion cell into an argon acceptor stream and then reacts with ozone in the flow-through chemiluminescence measuring cell of the flow system. Under optimal conditions, the method is characterized by a wide linear calibration range from 0.6 microg L(-1) to 25 mg L(-1), a detection limit of 0.6 microg L(-1) and a sample throughput of 300 samples per hour at 25 mg L(-1) and 450 samples per hour at 25 microg L(-1).
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Lab
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Atomic Emission Spectroscopy: Instrumentation