Related Experiment Video
Updated: Jul 16, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Validation of an analytical method to determine sulfamides in kidney by HPLC-DAD and PARAFAC2 with first-order
I García1, M C Ortiz, L Sarabia
1Department of Chemistry, Faculty of Sciences, Pza. Misael Bañuelos s/n, 09001 Burgos, Spain.
Abstract:
Six sulfamides were extracted from kidney and analysed by high-performance liquid chromatography with diode array detection (HPLC-DAD): sulfadiazine, sulfamethazine, sulfamethoxypyridazine, sulfamethoxazole, sulfadimethoxine and sulfaquinoxaline. Two main difficulties arose in identifying and quantifying the analytes. Firstly, the chromatographic peaks of the matrix interferences overlapped with those of the analytes. The uniqueness property of PARAFAC2 solved this problem. Secondly, the gradient elution caused a baseline drift. The first-order derivative of the chromatograms minimized its effect. The analytical method was validated. As the performance criteria detailed in the European Decision 2002/657/EC are based on specific signals, this paper generalizes those criteria for higher-order and non-specific signals. In this sense the proposed methodology is general and can be applied to any chromatographic method (HPLC or GC) with a detector that provide a multivariate signal (MS, DAD, EC, etc.).