Determination of output factors for small proton therapy fields
Jonas D Fontenot1, Wayne D Newhauser, Charles Bloch
1Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030, USA. jdfonteno@mdanderson.org
Abstract:
Current protocols for the measurement of proton dose focus on measurements under reference conditions; methods for measuring dose under patient-specific conditions have not been standardized. In particular, it is unclear whether dose in patient-specific fields can be determined more reliably with or without the presence of the patient-specific range compensator. The aim of this study was to quantitatively assess the reliability of two methods for measuring dose per monitor unit (DIMU) values for small-field treatment portals: one with the range compensator and one without the range compensator. A Monte Carlo model of the Proton Therapy Center-Houston double-scattering nozzle was created, and estimates of D/MU values were obtained from 14 simulated treatments of a simple geometric patient model. Field-specific D/MU calibration measurements were simulated with a dosimeter in a water phantom with and without the range compensator. D/MU values from the simulated calibration measurements were compared with D/MU values from the corresponding treatment simulation in the patient model. To evaluate the reliability of the calibration measurements, six metrics and four figures of merit were defined to characterize accuracy, uncertainty, the standard deviations of accuracy and uncertainty, worst agreement, and maximum uncertainty. Measuring D/MU without the range compensator provided superior results for five of the six metrics and for all four figures of merit. The two techniques yielded different results primarily because of high-dose gradient regions introduced into the water phantom when the range compensator was present. Estimated uncertainties (approximately 1 mm) in the position of the dosimeter in these regions resulted in large uncertainties and high variability in D/MU values. When the range compensator was absent, these gradients were minimized and D/MU values were less sensitive to dosimeter positioning errors. We conclude that measuring D/MU without the range compensator present provides more reliable results than measuring it with the range compensator in place.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Dose Size and Dosing Frequency: Determination Methods