Performance of the 24-2-5 frequency doubling technology screening test: a prospective case study
P G D Spry1, H M Hussin, J M Sparrow
1Bristol Eye Hospital, Lower Maudlin Street, Bristol, UK. paul.spry@ubht.nhs.uk
The frequency doubling technology (FDT) 24-2-5 screening test shows similar glaucoma detection ability to the N-30-5 test but is less reliable and takes longer. Both FDT screening tests demonstrate high discriminatory power for glaucoma detection.
Area of Science:
- Ophthalmology
- Optometry
- Medical Technology
Background:
- Glaucoma screening is crucial for early detection and prevention of vision loss.
- Frequency Doubling Technology (FDT) perimetry is a common method for visual field testing in glaucoma.
- Established FDT screening tests require evaluation against newer protocols.
Purpose of the Study:
- To compare the performance of the FDT 24-2-5 screening test against the N-30-5 FDT screening test.
- To assess discriminatory power, reliability, and test duration for glaucoma detection.
Main Methods:
- A prospective study involving individuals referred for possible glaucoma.
- Testing with FDT 24-2-5 and N-30-5 screening tests using Humphrey Matrix perimeter.
- Standard clinical examination and comparison of sensitivity, specificity, and reliability.
Main Results:
- Both tests showed high discriminatory power for glaucoma detection (AUC 0.87 for N-30-5, 0.92 for 24-2-5).
- The 24-2-5 test had higher sensitivity (83%) but lower specificity (75%) than the N-30-5 (78% sensitivity, 85% specificity).
- The 24-2-5 test was significantly longer (111s vs 39s) and yielded more unreliable results (36% vs 9%).
Conclusions:
- Minimal differences in discriminatory power were observed between the FDT 24-2-5 and N-30-5 screening tests.
- The N-30-5 test demonstrated better reliability and a shorter test duration compared to the 24-2-5.
- Both FDT screening tests are effective in detecting individuals with glaucoma.