Related Experiment Video
Updated: Jul 16, 2026

Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
Development of an HPLC method for the determination of doripenem in human and mouse serum
Christina Sutherland1, David P Nicolau
1Center for Anti-Infective Research and Development, Hartford Hospital, Hartford, CT 06102, United States.
Abstract:
A HPLC method utilizing solid phase extraction was developed to analyze doripenem (formerly S-4661) in human and mouse serum. A reversed-phase column was used with a UV detector set at 295 nm. The mobile phase consisted of methanol and phosphate buffer at a flow rate of 1.5 ml/min. Meropenem was used as the internal standard. The standard curve was linear over a range of 0.5-40 microg/ml. The assay is simple, reproducible, and accurate and has been used successfully to analyze doripenem concentrations from a murine pharmacokinetic study.