Related Experiment Video
Updated: Jul 16, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Ventilator assessment of respiratory mechanics in paediatric intensive care
Gopinathannair Harikumar1, Anne Greenough, Gerrard F Rafferty
1King's College London, MRC-Asthma Centre, Division of Asthma, Allergy and Lung Biology, Kings' College London School of Medicine at Guy's, King's College and St Thomas' Hospitals, London, UK.
Unlabelled:
Many modern "paediatric" mechanical ventilators have in-built features for estimation of respiratory mechanics which could be useful in the management of ventilated infants and children. The aim of this study was to determine if such measurements were reproducible and accurate. Ventilator (Draeger Evita 4) displayed compliance (Cvent) and resistance (Rvent) values were assessed and compared to the results of respiratory system mechanics (respiratory system compliance (Crs) and resistance (Rrs)) measurements obtained using a single breath occlusion technique. Seventeen children (median age 5.1; range 0.3 to 16 yrs) were studied on 24 occasions. The mean coefficients of variations for the techniques were similar (Cvent 13%; Crs 11%; Rvent 16%; Rrs 14%). The mean (SD) Crs (22.8 (12.3) ml/cmH2O) did not differ significantly from Cvent (22.1 (12.7) ml/cm H2O) but the mean Rrs 21.0 (12.7) cmH2O/l/s was significantly higher than the mean Rvent 32.0 (32.0) cmH2O/l/s (p = 0.03). Bland and Altman analysis demonstrated a mean difference of -10.94 cmH2O/l/s (SD 24.1) between Rrs and Rvent; the agreement between Rrs and Rvent decreased as Rrs increased (p = 0.008).
Conclusions:
Ventilator assessment of compliance, but not resistance, using the Evita 4 is reproducible and reliable.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)