Noninvasive liver-iron measurements with a room-temperature susceptometer
William F Avrin1, Sankaran Kumar
1Insight Magnetics, 9598 Carroll Canyon Road #165, San Diego, CA 92126, USA. bavrin@ucsd.edu
Abstract:
Magnetic susceptibility measurements on the liver can quantify iron overload accurately and noninvasively. However, established susceptometer designs, using superconducting quantum interference devices (SQUIDs) that work in liquid helium, have been too expensive for widespread use. This paper presents a less expensive liver susceptometer that works at room temperature. This system uses oscillating magnetic fields, which are produced and detected by copper coils. The coil design cancels the signal from the applied field, eliminating noise from fluctuations of the source-coil current and sensor gain. The coil unit moves toward and away from the patient at 1 Hz, cancelling drifts due to thermal expansion of the coils. Measurements on a water phantom indicated instrumental errors less than 30 microg of iron per gram of wet liver tissue, which is small compared with other errors due to the response of the patient's body. Liver-iron measurements on eight thalassemia patients yielded a correlation coefficient r = 0.98 between the room-temperature susceptometer and an existing SQUID. These results indicate that the fundamental accuracy limits of the room-temperature susceptometer are similar to those of the SQUID.