Related Experiment Video
Updated: Jul 15, 2026

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Development of a RP-HPLC method for screening potentially counterfeit anti-diabetic drugs
Jing Yao1, Ya-Qin Shi, Zhuo-Rong Li
1Institute of Medicinal Biotechnology, Chinese Academy of Medical Science and Peking Union Medical College, Beijing 100050, PR China. yjciae@gmail.com
Abstract:
Pharmaceutical counterfeiting is becoming a serious problem in the world, especially in developing countries including China. Herein an isocratic reversed-phase high performance liquid chromatography (RP-HPLC) method was developed for screening counterfeit medicines and adulterated dietary supplement products. The developed method could be employed to separate and determine simultaneously six anti-diabetic drugs (glipizide, gliclazide, glibenclamide, glimepiride, gliquidone, repaglinide) on an isocratic solvent system using an Alltima C18 column (5 microm, 150 mmx4.6 mm) with an isocratic mobile phase of methanol-phosphate buffer (pH 3.0; 0.01 mol/L) (70:30, v/v), at a flow rate of 1.0 mL/min and at a wavelength of 230 nm. The proposed method was successfully applied to the analysis of medicinal and dietary supplement samples purchased from the local market in China.