Effect of lower limb compression devices on thermodilution cardiac output measurement
Keith Killu1, John M Oropello, Anthony R Manasia
1Department of Surgery, The Mount Sinai School of Medicine, New York, NY, USA.
Objective:
The aim of this study was to determine whether lower limb (calf) sequential compression devices (SCDs) have a significant effect on thermodilution cardiac output measurements using a pulmonary artery catheter.
Design:
Prospective clinical investigation.
Setting:
Surgical and neurosurgical intensive care units in a university hospital.
Patients:
A total of 43 patients with pulmonary artery catheters and bilateral lower limb SCDs.
Measurements And Main Results:
Cardiac output was measured (average of three) when the SCDs were off (T1), during the first 2-4 secs of the inflation cycle (T2), during seconds 4-8 of the inflation cycle (T3), and when the SCDs were off again (T4). Cardiac output measurements were consistently lower when measured during the SCD inflation cycle. The decrease in cardiac output ranged from 7.58% to 49.5%, with a mean reduction of 24.51% in the first 2-4 seconds and 20.61% during seconds 4-8 (p < .001). Two patients displayed an increase in cardiac output during the inflation cycle; one patient had an increase of 2.78% and the other an increase of 13.5%. In 11 patients, measurements were also made using a pulse contour-analysis cardiac output device, but no changes in pulse contour-analysis cardiac output were observed during the same time period.
Conclusions:
Thermodilution cardiac output measurements via a pulmonary artery catheter should not be done during the inflation cycle of lower limb SCDs because they produce a falsely low cardiac output.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure in the Leg
Preparation:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.